₹44,500
₹21,800
₹25,000
₹25,000
₹44,500
₹21,800
₹25,000
₹25,000
Type | Radiant Warmers |
Material | Metal |
Model Name/Number | RHW A002 |
Brand | ACCULOC |
Color | NA |
Voltage | 220V AC |
Height | na |
Width | na |
Assay Time Min | na |
Temperature | Skin and Air Servo Controlled |
Length Mm | na |
Light Source | na |
Number Of Wheels | 4 |
Power V | 220VAC |
Temp Range Deg Celsius | 38 |
Weight | 55kgs |
Country of Origin | Made in India |
OPERATING MODES SKIN MODE: In Skin mode heater power is controlled with reference to Skin probe. In case when the temperature exceeds Max ( 38 °C or 101 °F) in any probe Heater is turned off. AIR MODE: In Air mode heater power is controlled with reference to Air probe. In case when the temperature exceeds Max ( 38 °C or 101 °F) in any probe Heater is turned off. MANUAL: MODE In manual mode user can control heater manually. In this mode user can adjust heating speed (0 to 8 speed) and set timer upto 59min. With probes installed when the temperature exceeds Max ( 38 °C or 101 °F) in any probe Heater is turned off. Special attention is needed while operating without
ALERTS Power fail: Power fail indicator turns ON with sound alert (cannot be muted). Probe fail: Probe fail indicator turns ON with sound alert (can be muted). Displays “-.-.-“on temp display. Heater overload: Heater overload indicator turns ON. Low temperature: Low alarm indicator turns ON with sound alert (can be muted). Displays blinks. High temperature: High alarm indicator turns ON with sound alert (can be muted). Displays blinks. Manual mode Time out: Short sound alert and timer flashing on LCD.
Type | Radiant Warmers |
Material | Metal |
Model Name/Number | RHW A002 |
Brand | ACCULOC |
Color | NA |
Voltage | 220V AC |
Height | na |
Width | na |
Assay Time Min | na |
Temperature | Skin and Air Servo Controlled |
Length Mm | na |
Light Source | na |
Number Of Wheels | 4 |
Power V | 220VAC |
Temp Range Deg Celsius | 38 |
Weight | 55kgs |
Country of Origin | Made in India |
OPERATING MODES SKIN MODE: In Skin mode heater power is controlled with reference to Skin probe. In case when the temperature exceeds Max ( 38 °C or 101 °F) in any probe Heater is turned off. AIR MODE: In Air mode heater power is controlled with reference to Air probe. In case when the temperature exceeds Max ( 38 °C or 101 °F) in any probe Heater is turned off. MANUAL: MODE In manual mode user can control heater manually. In this mode user can adjust heating speed (0 to 8 speed) and set timer upto 59min. With probes installed when the temperature exceeds Max ( 38 °C or 101 °F) in any probe Heater is turned off. Special attention is needed while operating without
ALERTS Power fail: Power fail indicator turns ON with sound alert (cannot be muted). Probe fail: Probe fail indicator turns ON with sound alert (can be muted). Displays “-.-.-“on temp display. Heater overload: Heater overload indicator turns ON. Low temperature: Low alarm indicator turns ON with sound alert (can be muted). Displays blinks. High temperature: High alarm indicator turns ON with sound alert (can be muted). Displays blinks. Manual mode Time out: Short sound alert and timer flashing on LCD.